1 Sonomicrometer Method  





2 Conductance catheter technique  





3 Theory  





4 Calibration and correction  





5 Parameters  





6 References  





7 See also  





8 External links  














Pressurevolume loop experiments







Add links
 









Article
Talk
 

















Read
Edit
View history
 








Tools
   


Actions  



Read
Edit
View history
 




General  



What links here
Related changes
Upload file
Special pages
Permanent link
Page information
Cite this page
Get shortened URL
Download QR code
Wikidata item
 




Print/export  



Download as PDF
Printable version
 




Print/export  



















Appearance
   

 






From Wikipedia, the free encyclopedia
 


This is an old revision of this page, as edited by 76.10.130.47 (talk)at18:00, 5 August 2010. The present address (URL) is a permanent link to this revision, which may differ significantly from the current revision.
(diff)  Previous revision | Latest revision (diff) | Newer revision  (diff)

Pressure Volume loops are widely used in basic and preclinical research. Left ventricular PV loops are considered to be the gold standard for hemodynamic assessment and are widely used in research to evaluate cardiac performance. While it has long been possible to measure pressure in real time from the left ventricle, measuring the volume was technically more difficult.

The use of ultrasonic sonomicrometry and the development of the conductance catheter triggered renewed interest in PV loops studies. In sonomicrometry, small ultrasonic transducers (usually referred to as "crystals") transmit signals to each other, and the distance between them is accurately determined based on the transit-time of the signals. By knowing the long and short axis lengths of the ventrical, ventricular volume is easily and accurately determined. Conductance cathethers measure instantaneous conductance in the left ventricle, which is then converted to blood volume using complex formulas and usually after determining and applying various correction factors. Typically only one method is used to perform PV studies in research settings.

The miniaturization of sonomicrometer crystals and pressure catheters have made mice PV loop studies feasible and more common.

Sonomicrometer Method

A sonomicrometer system is composed of an electronic signal-processing unit that is connected to small ultrasonic transducers (crystals). A computer acting as a data acquisition and display device obtains data in real time from the signal processor unit, while the crystals are implanted in or on the left ventrical. As few as 2 or as many as 6 crystals can be used to perform 1, 2 or 3-axis length measurements of the axial planes of the heart, usually from 200 to 2000 times per second. A typical sonomicrometer system has a resolution of 12 microns, enabling high-resolution measurements of the axial lengths.

Ventricular volume is computed directly (either in micro-liters or milli-liters) by combining the axial length measurements in standard spherical or ellipsoidal volume equations:

Volume = 4/3 x pi x r^3 (for a single-axis measurement) Volume = pi/6 x L1 x L2^2 (for a two-axis measurement, where L1 is the length of the long axis) Volume = pi/6 x L1 x L2 x L3 (for a three-axis measurement)

When the axial measurements are acquired in terms of mm, then the volume units in these equations will be in terms of milli-liters.


Conductance catheter technique

A conductance catheter contains two or more ring-shaped electrodes along its length. When a high-frequency low-amplitude constant current is passed through the outer electrodes to generate an electric field, the potential difference between any pair of inner electrodes is inversely proportional to the amount of conductive material at that site.

Conductance is defined as the applied current divided by the voltage measured between two adjacent electrodes.

The conductance catheter technique has no major drawbacks but requires careful calibration of conductance signals. Other techniques exist but this article focuses on the well-established conductance catheter technique.

(NOTE: For typical catheter configurations the linear relationship of the inverse potential difference to the amount of material is only an approximation. It is only valid for volumes where the diameter of that volume is less than the distance between the measuring electrodes.)

Theory

The formula by Baan et al. (1984) for obtaining ventricular volume is as follows: File:Wiki pv equation.jpg

where

The conductance measured by the catheter is actually the conductance of the blood and of the surrounding myocardial tissue; this latter conductance is called the parallel conductance (GP).

Calibration and correction

From the above equation, it can be seen that the volume measurement requires the knowledge of α, ρ, L and GP.

The equation may be considered to be a straight line of the form y = a + bx, where a (the gain) is a combination of the three terms 1/α,  and L2 and b (the offset) is GP.

L is available from catheter data sheets, while ρ can be measured directly using appropriate equipment.

Ideally, gain should be determined in every experiment. While L is known and ρ can be measured, the α factor poses more difficulties. Indeed, the only way to obtain a reliable gain (encompassing α) involves directly measuring cardiac output. Therefore, for most purposes, an educated estimate is used based on values found in the literature.

To take into account the parallel conductance, GP, the most common method involves injection of hypertonic saline into the subject - enough to temporarily reduce blood resistance, and therefore increase preload, but not so much as to alter haemodynamics. The minimum and maximum volumes (Vmax and Vmin) from each loop in the series of loops are plotted on a graph. Vmax and Vmin lines are extrapolated and at their point of intersection - where Vmax is Vmin, and therefore must be 0 - conductance is parallel conductance only. The volume at this point is the correction volume.

Calculation of correction volume

Parameters

Several parameters can be calculated for each loop, e.g. end-systolic pressure, end-systolic pressure, ejection and filling intervals, contractility index, stroke volume, ejection fraction etc.

More importantly, other interesting parameters are derived from series of loops obtained under changing conditions. For example, the end-diastolic pressure-volume relationship (EDPVR) and end-systolic pressure-volume relationship (ESPVR) are derived from series of loops obtained by slowly inflating a balloon to occlude the inferior vena cava - a procedure that reduces preload.

Calculation of EDPVR and ESPVR

EDPVR and ESPVR are valuable because they are load-independent indices of left ventricular function.

Other parameters derived from series of loops are:

References

Esposito G, Santana AL, Dilly K, Santos Cruz JD, Mao L, Lederer WJ, Rockman HA (2000) Cellular and functional defects in a mouse model of heart failure. Am J Physiol Heart Circ Physiol 279: H3101–H3112.

Sodums MT, Badke FR, Starling MR, Little WC, O'Rourke RA (1984) Evaluation of Left Ventricular Contratile Performance Utilizing End-Systolic Pressure-Volume Relationships in Conscious Dogs. Circulation Research. 54(6): 731-739.

Baan J et al. (1984). Continuous measurement of left ventricular volume in animals and humans by conductance catheter. Circulation 70: 812-823

Pacher P, Nagayama T, Mukhopadhyay P, Bátkai S, Kass DA (2008). Measurement of cardiac function using pressure–volume conductance catheter technique in mice and rats. Nature Protocols 3(9): 1422-34.

See also

External links

Sonometrics Corporation (Instrumentation, transducers and software for life science research labs)

emka TECHNOLOGIES SA Vendor of hardware and software for preclinical pv loop experiments

Video on Vimeo Video of pv-loops acquisition and analysis

Millar Instruments Inc Manufacturer of pressure-volume catheters

Scisense Manufacturers of hardware and pressure-volume catheters for small and large animal PV loop experiments


Retrieved from "https://en.wikipedia.org/w/index.php?title=Pressure–volume_loop_experiments&oldid=377338255"

Category: 
Cardiology
Hidden categories: 
Articles with multiple maintenance issues
Pages using multiple issues with unknown parameters
Articles with missing files
 



This page was last edited on 5 August 2010, at 18:00 (UTC).

This version of the page has been revised. Besides normal editing, the reason for revision may have been that this version contains factual inaccuracies, vandalism, or material not compatible with the Creative Commons Attribution-ShareAlike License.



Privacy policy

About Wikipedia

Disclaimers

Contact Wikipedia

Code of Conduct

Developers

Statistics

Cookie statement

Mobile view



Wikimedia Foundation
Powered by MediaWiki