1 Total ion current (TIC) chromatogram  





2 Base peak chromatogram  





3 Extracted-ion chromatogram (EIC or XIC)  





4 Selected-ion monitoring chromatogram (SIM)  





5 Selected-reaction monitoring chromatogram (SRM, MRM)  





6 See also  





7 References  














Mass chromatogram






Italiano
Türkçe

 

Edit links
 









Article
Talk
 

















Read
Edit
View history
 








Tools
   


Actions  



Read
Edit
View history
 




General  



What links here
Related changes
Upload file
Special pages
Permanent link
Page information
Cite this page
Get shortened URL
Download QR code
Wikidata item
 




Print/export  



Download as PDF
Printable version
 
















Appearance
   

 






From Wikipedia, the free encyclopedia
 

(Redirected from Total ion current)

Amass chromatogram is a representation of mass spectrometry data as a chromatogram, where the x-axis represents time and the y-axis represents signal intensity.[1] The source data contains mass information; however, it is not graphically represented in a mass chromatogram in favor of visualizing signal intensity versus time. The most common use of this data representation is when mass spectrometry is used in conjunction with some form of chromatography, such as in liquid chromatography–mass spectrometryorgas chromatography–mass spectrometry. In this case, the x-axis represents retention time, analogous to any other chromatogram. The y-axis represents signal intensity or relative signal intensity. There are many different types of metrics that this intensity may represent, depending on what information is extracted from each mass spectrum.

Total ion current (TIC) chromatogram[edit]

An example total ion current chromatogram from an LC-MS analysis.

The total ion current (TIC) chromatogram represents the summed intensity across the entire range of masses being detected at every point in the analysis.[2][3] The range is typically several hundred mass-to-charge units or more. In complex samples, the TIC chromatogram often provides limited information as multiple analytes elute simultaneously, obscuring individual species.

Base peak chromatogram[edit]

An example base peak chromatogram from an LC-MS analysis.

The base peak chromatogram is similar to the TIC chromatogram, however it monitors only the most intense peak in each spectrum.[4][3] This means that the base peak chromatogram represents the intensity of the most intense peak at every point in the analysis. Base peak chromatograms often have a cleaner look and thus are more informative than TIC chromatograms because the background is reduced by focusing on a single analyte at every point.

Extracted-ion chromatogram (EIC or XIC)[edit]

An example extracted-ion chromatogram from an LC-MS analysis.

In an extracted-ion chromatogram (EIC or XIC), also called a reconstructed-ion chromatogram (RIC), one or more m/z values representing one or more analytes of interest are recovered ('extracted') from the entire data set for a chromatographic run.[3] The total intensity or base peak intensity within a mass tolerance window around a particular analyte's mass-to-charge ratio is plotted at every point in the analysis. The size of the mass tolerance window typically depends on the mass accuracy and mass resolution of the instrument collecting the data. This is useful for re-examining data to detect previously-unsuspected analytes, to highlight potential isomers, resolve suspected co-eluting substances, or to provide clean chromatograms of compounds of interest. An extracted-ion chromatogram is generated by separating the ions of interest from a data file containing the full mass spectrum over time after the fact; this is different from selected-ion chromatograms, discussed below, in which data is collected only for specific m/z values. A closely related term is extracted-compound chromatogram (ECC).

Selected-ion monitoring chromatogram (SIM)[edit]

Aselected-ion monitoring (SIM) chromatogram is similar to an EIC/XIC, with the exception that the mass spectrometer is operated in SIM mode, such that only preselected m/z values are detected in the analysis.[5] SIM experiments can be performed using mass spectrometry (MS) or tandem mass spectrometry (MS/MS) instruments. They are more common on MS instruments. This differs significantly from the extracted-ion chromatogram mentioned above in that only data for the ion(s) of interest are collected in a SIM experiment; for extracted-ion chromatograms (EIC or XIC), data for an entire mass range are collected during the run and then examined for analytes of interest after the completion of the run.

Selected-reaction monitoring chromatogram (SRM, MRM)[edit]

The selected-reaction monitoring (SRM) experiment is very similar to the SIM experiment except that tandem mass spectrometry is used and a specific product ion of a specific parent ion is detected.[3] The mass of the parent analyte is first selected while other ions are filtered away. The parent analyte ion is then fragmented in the gas phase and a specific fragment ion is monitored. This experiment has very high specificity because the SRM chromatogram represents only ions of a particular mass that fragment in a manner that produce a very specific product mass. This type of experiment can only be performed using tandem mass spectrometry. The technology progress in the MS/MS area lead to the development of MRM, Multiple Reaction Monitoring, which allows simultaneous detection of several coeluting analytes with different parent and/or product ions.

See also[edit]

References[edit]

  1. ^ Hites, Ronald A.; Biemann, Klaus (July 1970). "Computer evaluation of continuously scanned mass spectra of gas chromatographic effluents". Analytical Chemistry. 42 (8): 855–860. doi:10.1021/ac60290a009.
  • ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "total ion current". doi:10.1351/goldbook.T06410
  • ^ a b c d Murray, Kermit K.; Boyd, Robert K.; Eberlin, Marcos N.; Langley, G. John; Li, Liang; Naito, Yasuhide (2013). "Definitions of terms relating to mass spectrometry (IUPAC Recommendations 2013)". Pure and Applied Chemistry. 85 (7): 1515–1609. doi:10.1351/PAC-REC-06-04-06. ISSN 0033-4545.
  • ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "base peak". doi:10.1351/goldbook.B00608
  • ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "selected ion monitoring". doi:10.1351/goldbook.S05547

  • Retrieved from "https://en.wikipedia.org/w/index.php?title=Mass_chromatogram&oldid=1127607435#Total_ion_current_(TIC)_chromatogram"

    Category: 
    Mass spectrometry
     



    This page was last edited on 15 December 2022, at 17:50 (UTC).

    Text is available under the Creative Commons Attribution-ShareAlike License 4.0; additional terms may apply. By using this site, you agree to the Terms of Use and Privacy Policy. Wikipedia® is a registered trademark of the Wikimedia Foundation, Inc., a non-profit organization.



    Privacy policy

    About Wikipedia

    Disclaimers

    Contact Wikipedia

    Code of Conduct

    Developers

    Statistics

    Cookie statement

    Mobile view



    Wikimedia Foundation
    Powered by MediaWiki