Analytical thermal desorption







Add links
 









Article
Talk
 

















Read
Edit
View history
 








Tools
   


Actions  



Read
Edit
View history
 




General  



What links here
Related changes
Upload file
Special pages
Permanent link
Page information
Cite this page
Get shortened URL
Download QR code
Wikidata item
 




Print/export  



Download as PDF
Printable version
 
















Appearance
   

 






From Wikipedia, the free encyclopedia
 


Analytical thermal desorption, known within the analytical chemistry community simply as "thermal desorption" (TD), is a technique that concentrates volatile organic compounds (VOCs) in gas streams prior to injection into a gas chromatograph (GC). It can be used to lower the detection limits of GC methods, and can improve chromatographic performance by reducing peak widths.[1]

History

[edit]

Analytical thermal desorption originated in the mid-1970s as an adaptation to the injection procedure for GC. Injector liners were packed with a compound able to adsorb organic compounds, used to sample air or gas, and then dropped into the inlet of the GC. This principle was first widely employed for occupational monitoring, in the form of personal badge-type monitors containing a removable charcoal strip.[2] These offered the advantage of being amenable to analysis without a separate solvent-extraction step.

Also developed in the 1970s was a method by which volatiles in the air were collected by diffusion onto tubes packed with a sorbent, which was then heated to release the volatiles into the GC system. These were first introduced for monitoring sulfur dioxide[3] and nitrogen dioxide,[4] but the analyte scope later widened as the sorbents became more advanced. Another early method (closely related to the modern purge-and-trap procedure) involved passing a stream of gas through a water sample to release the volatiles, which were again collected on a sorbent-packed tube.[5]

Such axial-type samplers, which later became known as 'sorbent tubes', were laid out as an industry standard in the late 1970s, by Working Group 5 (WG5) of the UK Health & Safety's Committee on Analytical Requirements (HSE CAR). The tubes they outlined were 3+12 inches long with an outer diameter of 14 inch, and were first employed in Perkin Elmer's ATD-50 instrument.[6]

At the same time, WG5 specified various basic functionality requirements for thermal desorption, and in the years since then, a number of improvements have been made to instrumentation for thermal desorption, including two-stage operation (see below), splitting and re-collection of samples, improved trap-cooling technology, standard system checks, and automation.

Principles

[edit]

Thermal desorption fundamentally involves collecting volatile organic compounds onto a sorbent, and then heating this sorbent in a flow of gas to release the compounds and concentrate them into a smaller volume.

Early thermal desorbers used just single-stage operation, whereby the volatiles collected on a sorbent tube were released by heating the tube in a flow of gas, from where they passed directly into the GC.

Modern thermal desorbers can also accommodate two-stage operation, whereby the gas stream from the sorbent tube (typically 100–200 mL) is collected on a narrower tube integral to the thermal desorber, called the focusing trap or cold trap. Heating this trap releases the analytes once again, but this time in an even smaller volume of gas (typically 100–200 μL), resulting in improved sensitivity and better GC peak shape.[1]

Modern thermal desorbers can accommodate both single-stage and two-stage operation, although single-stage operation is now usually carried out using the focusing trap to collect the analytes, rather than a sorbent tube.

It is normal for the focusing trap to be held at or below room temperature, although a temperature no lower than 0 °C is sufficient for all but the most volatile analytes. Higher trap temperatures also reduce the amount of water condensing inside the trap (when transferred to the GC column, water can reduce the quality of the chromatography).

Sampling configurations

[edit]

A wide variety of sampling configurations are used for thermal desorption, depending on the application. The most popular are listed below.

Single-stage thermal desorption

[edit]

This involves sampling direct onto the focusing trap of the thermal desorber. It is generally used for situations where the analytes are too volatile to be retained on sorbent tubes.

Two-stage thermal desorption

[edit]

This involves sampling first onto a sorbent tube. The most widely used tubes are those following the pattern laid out by WG5 (see above). After sampling (for which a variety of accessories are available), the tube is desorbed to transfer the analytes to the focusing trap before the second desorption stage transfers them to the GC. The greater sensitivity of this method has made it increasingly popular for sampling dilute gas streams, or in exploratory work where the target atmosphere is unknown.

Sorbents

[edit]

The sorbent tube and the focusing trap may be packed with one or more sorbents. The type and number of sorbents depends on a number of factors including the sampling setup, the analyte volatility range, analyte concentration, and the humidity of the sample.[9][10]

One of the most versatile and popular sorbents for thermal desorption is poly(2,6-diphenyl-p-phenylene oxide), known by its trademark Tenax.[11]

Analyte range

[edit]

Depending upon the sampling technique and the analytical conditions, thermal desorption can be used to reliably sample analytes ranging in volatility from ethane to about tetracontane (n-C40H82). Incompatible compounds include:

Applications

[edit]

Applications of thermal desorption were originally restricted to occupational health monitoring, but have since extended to cover a much wider range. Some of the most important are mentioned below – where available, examples of early reports, and more recent citations (including those of widely used standard methods) have been given:

References

[edit]
  1. ^ a b E. Woolfenden, Thermal desorption for gas chromatography, in: Gas Chromatography, ed. C.F. Poole, Elsevier, 2012, chapter 10, pp. 235–289; Analytical thermal desorption: History, technical aspects and application range, Thermal Desorption Technical Support Note 12, Markes International, April 2012, http://www.markes.com/Downloads/Application-notes.aspx.
  • ^ a b Lautenberger, W.J.; Kring, E.V.; Morello, J.A. (1980). "A new personal badge monitor for organic vapors". American Industrial Hygiene Association Journal. 1980 (41): 737–747. doi:10.1080/15298668091425581. PMID 7435378.
  • ^ Palmes, E.D.; Gunnison, A.F. (1973). "Personal monitoring device for gaseous contaminants". American Industrial Hygiene Association Journal. 34 (2): 78–81. doi:10.1080/0002889738506810. PMID 4197577.
  • ^ Palmes, E.D.; Gunnison, A.F.; DiMattio, J.; Tomczyk, C. (1976). "Personal sampler for nitrogen dioxide". American Industrial Hygiene Association Journal. 37 (10): 570–577. doi:10.1080/0002889768507522. PMID 983946.
  • ^ Badings, H.T.; Cooper, R.P.M. (1985). "Automatic system for rapid analysis of volatile compounds by purge-and-cold-trapping/capillary gas chromatography". Journal of High Resolution Chromatography and Chromatography Communications. 8 (11): 755–763. doi:10.1002/jhrc.1240081111.
  • ^ J. Kristensson, The use of ATD-50 system with fused silica capillaries in dynamic headspace analysis, in: Analysis of volatiles, ed. P. Schreier, De Gruyter, 1984, pp. 109-120.
  • ^ Vas, G.; Vékey, K. (2004). "Solid-phase microextraction: a powerful sample preparation tool prior to mass spectrometric analysis". Journal of Mass Spectrometry. 2004 (39): 233–254. doi:10.1002/jms.606. PMID 15039931.
  • ^ A.-L. Sunesson, Passive sampling in combination with thermal desorption and gas chromatography as a tool for assessment of chemical exposure, in: Comprehensive Analytical Chemistry, Volume 48: Passive Sampling Techniques in Environmental Monitoring, ed. R. Greenwood, G. Mills and B. Vrana, Elsevier, 2007.
  • ^ Woolfenden, E. (2010). "Sorbent-based sampling methods for volatile and semi-volatile organic compounds in air. Part 1: Sorbent-based air monitoring options". Journal of Chromatography A. 1217 (16): 2674–2684. doi:10.1016/j.chroma.2009.12.042. PMID 20106481.
  • ^ Woolfenden, E. (2010). "Sorbent-based sampling methods for volatile and semi-volatile organic compounds in air. Part 2. Sorbent selection and other aspects of optimizing air monitoring methods". Journal of Chromatography A. 1217 (16): 2685–2694. doi:10.1016/j.chroma.2010.01.015. PMID 20106482.
  • ^ Zlatkis, A.; Lichtenstein, A.; Tishbee, A. (1973). "Concentration and analysis of trace volatile organics in gases and biological fluids with a new solid adsorbent". Chromatographia. 6 (2): 67–70. doi:10.1007/BF02270540. S2CID 95423469.
  • ^ Bruner, F.; Ciccioli, P.; Nardo, F. Di (1974). "Use of graphitized carbon black in environmental analysis". Journal of Chromatography. 99: 661–672. doi:10.1016/s0021-9673(00)90893-8. PMID 4422759.
  • ^ Pankow, J.F.; Isabelle, L.M.; Hewetson, J.P.; Cherry, J.A. (1984). "A syringe and cartridge method for down-hole sampling for trace organics in ground water". Ground Water. 22 (3): 330–339. doi:10.1111/j.1745-6584.1984.tb01405.x.
  • ^ US EPA Compendium Method TO-17: Determination of volatile organic compounds in ambient air using active sampling onto sorbent tubes, US Environmental Protection Agency, January 1999, PDF
  • ^ W.R. Betz, S.G. Maroldo, G.D. Wachob and M.C. Firth, Characterization of carbon molecular sieves and activated charcoal for use in airborne contaminant sampling, American Industrial Hygiene Association Journal, 1989, 50: 181–187.
  • ^ Houldsworth, H.B.; O'Sullivan, J.; Musgrave, N. (1982). "Passive monitors for the determination of personal nitrous oxide exposure levels". Anaesthesia. 37 (4): 467–468. doi:10.1111/j.1365-2044.1982.tb01175.x. PMID 7081695. S2CID 37132681.
  • ^ MDHS 80, Laboratory method using diffusive solid sorbent tubes, thermal desorption and gas chromatography, UK Health & Safety Executive, August 1995, PDF
  • ^ Grote, A.A.; Kennedy, E.R. "Workplace monitoring for VOCs using thermal desorption-GC-MS". Journal of Environmental Monitoring. 2002 (4): 679–684.
  • ^ E. Woolfenden, Standardized methods for testing emissions of organic vapors from building products to indoor air, in: Organic Indoor Air Pollutants (2nd edn), ed. T. Salthammer and E. Uhde, Wiley-VCH, 2009, chapter 6, http://eu.wiley.com/WileyCDA/WileyTitle/productCd-3527312676.html.
  • ^ Method VDA 278: Thermal Desorption Analysis of Organic Emissions for the Characterization of Non-Metallic Materials for Automobiles, October 2011, http://www.vda.de/en/publikationen/publikationen_downloads/detail.php?id=1027.
  • ^ Kessler, A.; Baldwin, I.T. (2001). "Defensive function of herbivore-induced plant volatile emissions in nature". Science. 291 (5511): 2141–2144. doi:10.1126/science.291.5511.2141. PMID 11251117.
  • ^ Manolis, A. (1983). "The diagnostic potential of breath analysis". Clinical Chemistry. 29: 5–15. doi:10.1093/clinchem/29.1.5. PMID 6336681.
  • ^ E. Woolfenden, Flavour and fragrance profiling by ATD/GC, Laboratory Equipment Digest, April 1989, pp. 23–25.
  • ^ Kelly, L.; Woolfenden, E.A. "Enhanced GC-MS aroma profiling using thermal desorption technologies". Separation Science. 2008 (1): 16–23.
  • ^ Workplace chemical monitoring: Monitoring considerations, in: Occupational Health and Workplace Monitoring at Chemical Agent Disposal Facilities, Board on Army Science and Technology (National Research Council), 2001, chapter 2.

  • Retrieved from "https://en.wikipedia.org/w/index.php?title=Analytical_thermal_desorption&oldid=1225882139"

    Category: 
    Analytical chemistry
    Hidden categories: 
    Articles with short description
    Short description matches Wikidata
     



    This page was last edited on 27 May 2024, at 08:29 (UTC).

    Text is available under the Creative Commons Attribution-ShareAlike License 4.0; additional terms may apply. By using this site, you agree to the Terms of Use and Privacy Policy. Wikipedia® is a registered trademark of the Wikimedia Foundation, Inc., a non-profit organization.



    Privacy policy

    About Wikipedia

    Disclaimers

    Contact Wikipedia

    Code of Conduct

    Developers

    Statistics

    Cookie statement

    Mobile view



    Wikimedia Foundation
    Powered by MediaWiki