(Top)
 


1 Principle  



1.1  Stimulation  







2 References  














Impulse oscillometry







Add links
 









Article
Talk
 

















Read
Edit
View history
 








Tools
   


Actions  



Read
Edit
View history
 




General  



What links here
Related changes
Upload file
Special pages
Permanent link
Page information
Cite this page
Get shortened URL
Download QR code
 




Print/export  



Download as PDF
Printable version
 
















Appearance
   

 






From Wikipedia, the free encyclopedia
 


Impulse oscillometry (IOS), also known as respiratory oscillometry, forced oscillatory technique (FOT), or just oscillometry, is a non-invasive lung function test that measures the mechanical properties of the respiratory system, particularly the upper and intrathoracic airways, lung tissue and chest wall, usually during the patient's tidal breathing (the way someone breathes when they are relaxed).[1][2][3]

Principle[edit]

Impulse oscillometry measures the mechanical impedance of the respiratory system (Zrs), which encompasses the resistance of the respiratory system to flow (Rrs), the reactance or stiffness of the lung parenchyma in response to changes in volume (Xrs) and the inertance of accelerating gas in the airways (Irs).[2][3]

The following relations hold between these parameters: , where is the imaginary unit (), and , where is the airway elastance and is the angular velocity such that , where is the frequency of the stimulus oscillation.[2]

Zrs is measured by comparing the magnitudes of mechanical stimuli, specifically oscillations of pressure, i.e. pressure waves, transmitted into the respiratory system with the magnitudes of the stimuli's effects on tidal airflow; this is done by superimposing these oscillations over spontaneous tidal breathing.[1][2][3]

Stimulation[edit]

The stimulus is an oscillation of pressure of a particular frequency that is transmitted to the lungs of the patient.[1] This is usually done by mouth, though the direct stimulation of the chest wall is also possible.[2] These pressure waves cause changes in the airflow during tidal breathing; the magnitudes of the pressure waves and the changes in airflow they cause are then used to determine the airways' mechanical impedance.[1]

Frequencies ranging from 4-50 Hz are commonly generated by a loudspeaker, while frequencies between 0.5 and 4 Hz may alternatively also be generated by a pistonorpneumatic proportional solenoid valves.[2]

Different frequencies measure the mechanical properties of different parts of the respiratory system; the resistance at 5 Hz (R5) represents total airway resistance, while the resistance at 20 Hz (R20) represents the resistance of the central airways. The reactance at 5 Hz (X5) reflects the elasticity of the peripheral airways.[4]

References[edit]

  1. ^ a b c d Desiraju, Koundinya; Agrawal, Anurag (2016). "Impulse oscillometry: The state-of-art for lung function testing". Lung India. 33 (4): 410. doi:10.4103/0970-2113.184875. ISSN 0970-2113. PMC 4948229. PMID 27578934.
  • ^ a b c d e f King, Gregory G.; Bates, Jason; Berger, Kenneth I.; Calverley, Peter; Melo, Pedro L. de; Dellacà, Raffaele L.; Farré, Ramon; Hall, Graham L.; Ioan, Iulia; Irvin, Charles G.; Kaczka, David W.; Kaminsky, David A.; Kurosawa, Hajime; Lombardi, Enrico; Maksym, Geoffrey N. (2020-02-01). "Technical standards for respiratory oscillometry". European Respiratory Journal. 55 (2). doi:10.1183/13993003.00753-2019. hdl:10067/1678010151162165141. ISSN 0903-1936. PMID 31772002.
  • ^ a b c Kaminsky, David A.; Simpson, Shannon J.; Berger, Kenneth I.; Calverley, Peter; Melo, Pedro L. de; Dandurand, Ronald; Dellacà, Raffaele L.; Farah, Claude S.; Farré, Ramon; Hall, Graham L.; Ioan, Iulia; Irvin, Charles G.; Kaczka, David W.; King, Gregory G.; Kurosawa, Hajime (2022-03-31). "Clinical significance and applications of oscillometry". European Respiratory Review. 31 (163). doi:10.1183/16000617.0208-2021. hdl:2445/201561. ISSN 0905-9180. PMID 35140105.
  • ^ Qvarnström, Björn; Engström, Gunnar; Frantz, Sophia; Zhou, Xingwu; Zaigham, Suneela; Sundström, Johan; Janson, Christer; Wollmer, Per; Malinovschi, Andrei (2023-09-25). "Impulse oscillometry indices in relation to respiratory symptoms and spirometry in the Swedish Cardiopulmonary Bioimage Study". ERJ Open Research. 9 (5): 00736–2022. doi:10.1183/23120541.00736-2022. ISSN 2312-0541. PMC 10518858. PMID 37753278.

  • Retrieved from "https://en.wikipedia.org/w/index.php?title=Impulse_oscillometry&oldid=1227027686"

    Categories: 
    Pulmonary function testing
    Respiratory therapy
    Respiratory physiology
    Hidden categories: 
    Articles with short description
    Short description with empty Wikidata description
     



    This page was last edited on 3 June 2024, at 06:57 (UTC).

    Text is available under the Creative Commons Attribution-ShareAlike License 4.0; additional terms may apply. By using this site, you agree to the Terms of Use and Privacy Policy. Wikipedia® is a registered trademark of the Wikimedia Foundation, Inc., a non-profit organization.



    Privacy policy

    About Wikipedia

    Disclaimers

    Contact Wikipedia

    Code of Conduct

    Developers

    Statistics

    Cookie statement

    Mobile view



    Wikimedia Foundation
    Powered by MediaWiki