1 Transvenous ICD (leads in the heart)  



1.1  Pros  





1.2  Cons  





1.3  Patient selection  







2 Subcutaneous ICD (lead under the skin)  



2.1  Pros  





2.2  Cons  





2.3  Patient selection  







3 Transvenous vs subcutaneous ICD implantation procedure  





4 References  





5 External links  














Subcutaneous implantable defibrillator






Deutsch
Italiano
Nederlands
 

Edit links
 









Article
Talk
 

















Read
Edit
View history
 








Tools
   


Actions  



Read
Edit
View history
 




General  



What links here
Related changes
Upload file
Special pages
Permanent link
Page information
Cite this page
Get shortened URL
Download QR code
Wikidata item
 




Print/export  



Download as PDF
Printable version
 




In other projects  



Wikimedia Commons
 
















Appearance
   

 






From Wikipedia, the free encyclopedia
 

(Redirected from S-ICD)

S-ICD lead and generator position

Subcutaneous implantable cardioverter defibrillator, or S-ICD, is an implantable medical device for detecting and terminating ventricular tachycardia and ventricular fibrillation in patients at risk of sudden cardiac arrest.[1] It is a type of implantable cardioverter defibrillator but unlike the transvenous ICD, the S-ICD lead is placed just under the skin, leaving the heart and veins untouched.

The S-ICD was developed to reduce the risk of complications associated with transvenous leads.[2] Potential complications, such as infections in the bloodstream and the need to remove or replace the leads in the heart, are minimised or entirely eliminated with the S-ICD system.

Transvenous ICD (leads in the heart)[edit]

Pros[edit]

The generator is smaller than the S-ICD generator, which may result in a less visible implanted device. This could improve the time needed to get used to the implantable device, although this is subjective. The procedure can usually be done under local anesthesia and light sedation. The transvenous ICD is capable of pacing for bradycardia and delivering antitachycardia pacing (ATP). However, device-related complications were numerically more frequent in patients with transvenous ICDs, inappropriate shocks are less frequent that in those with subcutaneous ICDs.[3]

Cons[edit]

The leads go into the vein and heart and will grow into the heart wall over time. This may increase the chance of complications if the leads need to be removed or replaced, as the procedure to extract an intracardiac leads can be a challenge. Because the leads need to go into the heart they need to be relatively thin and flexible, since they have to pass through (and remain in) the heart valve(s) and need to flex with every heartbeat. This makes the leads more vulnerable to lead fracture (and therefore complications). It has been demonstrated that device-related complications were numerically more frequent in patients with transvenous ICDs.[3] Due to the position of the pulse generator under the collarbone, it can be more visible with clothing with low neckline.

Patient selection[edit]

Patients who are relatively older, who need ICD for secondary prevention, or who have concomitant bradycardia requiring pacing, or heart failure requiring cardiac resynchronisation therapy are more suitable for transvenous ICD implantation. An older patient with ischemic cardiomyopathy and documented symptomatic ventricular tachycardia is a typical example.[4]

Subcutaneous ICD (lead under the skin)[edit]

Pros[edit]

The lead does not go into the heart, which means it leaves the veins and the heart completely intact. This reduces chance of complications (e.g. systemic infections). Because the lead does not go into the heart it can be thicker and more robust. This minimizes / reduces the chance of lead fracture. In the event the system needs to be explanted, the procedure is a relatively simple surgical procedure.

Cons[edit]

The pulse generator is larger than most transvenous ICD pulse generators. This could result in a longer time needed to get used to it, although this is subjective. Depending on the physique of a person, the S-ICD may be more visible with bare chest. The procedure usually requires deep sedation or general anaesthesia, as creating a larger pocket between the muscles and tunnelling the lead over the sternum, as well as performing defibrillation threshold testing, can be quite painful. The S-ICD can deliver only temporary post-shock pacing, but cannot otherwise address bradycardia and cannot deliver anti-tachycardia pacing. Inappropriate shocks were numerically more frequent in those with subcutaneous ICDs.[3] Defibrillation testing has traditionally been considered mandatory in patients with subcutaneous implantable cardioverter–defibrillator to confirm appropriate ventricular fibrillation detection.[5] However, PRAETORIAN-DFT randomised clinical trial is aiming to demonstrate non-inferiority of omitting DFT in patients undergoing S-ICD implantation in which the S-ICD system components are optimally positioned by calculated PRAETORIAN score.[6]

Patient selection[edit]

Patients who are relatively younger, who need ICD for primary prevention, and who do not require pacing or cardiac resynchronisation therapy, are more suitable for S-ICD implantation. A young survivor of aborted sudden cardiac death is a typical example.

Transvenous vs subcutaneous ICD implantation procedure[edit]

Transvenous ICD implant procedure Subcutaneous ICD implant procedure
A transvenous ICD is typically implanted in the left shoulder area, near the collarbone. Occasionally the right side is preferred for certain patients or other specific reasons. In contrast to a transvenous ICD, the pulse generator is implanted on the left side of the chest next to the rib cage, just under the arm, and the lead is implanted just under the skin above the breastbone.
Using X-ray imaging (fluoroscopy), the leads are fed through a vein into the heart and through the heart valve(s) into the heart. Guided by anatomical landmarks and/or an X-ray image, the subcutaneous ICD electrode is tunneled under the skin. The subcutaneous ICD delivers therapy without the need for wires implanted in the heart.
Depending on heart condition, 1, 2 or 3 leads will be placed in the heart. Once the leads are put in place, they are attached to the heart wall for optimal connectivity. The subcutaneous ICD leaves the heart and blood vessels untouched and intact.

References[edit]

  1. ^ Westerman, Stacy B; El-Chami, Mikhael (2018). "The subcutaneous implantable cardioverter defibrillator––review of the recent data". Journal of Geriatric Cardiology. 15 (3): 222–228. doi:10.11909/j.issn.1671-5411.2018.03.004 (inactive 2024-05-10). ISSN 1671-5411. PMC 5919810. PMID 29721001.{{cite journal}}: CS1 maint: DOI inactive as of May 2024 (link)
  • ^ Baalman, S. W. E.; Quast, A. B. E.; Brouwer, T. F.; Knops, R. E. (2018). "An Overview of Clinical Outcomes in Transvenous and Subcutaneous ICD Patients". Current Cardiology Reports. 20 (9): 72. doi:10.1007/s11886-018-1021-8. ISSN 1523-3782. PMC 6061190. PMID 29992422.
  • ^ Poole, Jeanne E.; Olshansky, Brian; Mark, Daniel B.; Anderson, Jill; Johnson, George; Hellkamp, Anne S.; Davidson-Ray, Linda; Fishbein, Daniel P.; Boineau, Robin E.; Anstrom, Kevin J.; Reinhall, Per G.; Packer, Douglas L.; Lee, Kerry L.; Bardy, Gust H. (2020-07-28). "Long-Term Outcomes of Implantable Cardioverter-Defibrillator Therapy in the SCD-HeFT". Journal of the American College of Cardiology. 76 (4): 405–415. doi:10.1016/j.jacc.2020.05.061. ISSN 0735-1097. PMID 32703511.
  • ^ Waroux, Jean-Benoit le Polain de; Ploux, Sylvain; Mondoly, Pierre; Eschalier, Romain; Strik, Marc; Houard, Laura; Pierre, Bertrand; Buliard, Samuel; Klotz, Nicolas; Ritter, Philippe; Haissaguerre, Michel (2018-05-01). "Defibrillation testing is mandatory in patients with subcutaneous implantable cardioverter–defibrillator to confirm appropriate ventricular fibrillation detection". Heart Rhythm. 15 (5): 642–650. doi:10.1016/j.hrthm.2018.02.013. ISSN 1547-5271. PMID 29709229.
  • ^ Quast, Anne-Floor B. E.; Baalman, Sarah W. E.; Betts, Tim R.; Boersma, Lucas V. A.; Bonnemeier, Hendrik; Boveda, Serge; Brouwer, Tom F.; Burke, Martin C.; Delnoy, Peter Paul H. M.; El-Chami, Mikhael; Kuschyk, Juergen (August 2019). "Rationale and design of the PRAETORIAN-DFT trial: A prospective randomized CompArative trial of SubcutanEous ImplanTable CardiOverter-DefibrillatoR ImplANtation with and without DeFibrillation testing" (PDF). American Heart Journal. 214: 167–174. doi:10.1016/j.ahj.2019.05.002. ISSN 1097-6744. PMID 31220775. S2CID 181898926.
  • External links[edit]


    Retrieved from "https://en.wikipedia.org/w/index.php?title=Subcutaneous_implantable_defibrillator&oldid=1223591805"

    Categories: 
    Cardiology
    Heart
    Cardiac electrophysiology
    Implants (medicine)
    Medical devices
    Cardiac procedures
    Hidden categories: 
    CS1 maint: DOI inactive as of May 2024
    Articles needing additional references from March 2016
    All articles needing additional references
    Wikipedia articles with style issues from March 2016
    All articles with style issues
    Articles with multiple maintenance issues
     



    This page was last edited on 13 May 2024, at 04:12 (UTC).