Linear sweep voltammetry






Español
Italiano
 

Edit links
 









Article
Talk
 

















Read
Edit
View history
 








Tools
   


Actions  



Read
Edit
View history
 




General  



What links here
Related changes
Upload file
Special pages
Permanent link
Page information
Cite this page
Get shortened URL
Download QR code
Wikidata item
 




Print/export  



Download as PDF
Printable version
 
















Appearance
   

 






From Wikipedia, the free encyclopedia
 


Linear potential sweep

Inanalytical chemistry, linear sweep voltammetry is a method of voltammetry where the current at a working electrode is measured while the potential between the working electrode and a reference electrode is swept linearly in time.[1][2] Oxidation or reduction of species is registered as a peak or trough in the current signal at the potential at which the species begins to be oxidized or reduced.

Experimental method

[edit]
Comparison of the current response of a platinum disc electrode in 1 M sulphuric acid given by linear sweep cyclic voltammetry and staircase cyclic voltammetry methods. Staircase voltammetry suppresses the non-faradaic adsorption of hydrogen.

The experimental setup for linear sweep voltammetry utilizes a potentiostat and a three-electrode setup to deliver a potential to a solution and monitor its change in current. The three-electrode setup consists of a working electrode, an auxiliary electrode, and a reference electrode. The potentiostat delivers the potentials through the three-electrode setup. A potential, E, is delivered through the working electrode. The slope of the potential vs. time graph is called the scan rate and can range from mV/s to 1,000,000 V/s.[3] The working electrode is one of the electrodes at which the oxidation/reduction reactions occur—the processes that occur at this electrode are the ones being monitored. The auxiliary electrode (or counter electrode) is the one at which a process opposite from the one taking place at the working electrode occurs. The processes at this electrode are not monitored. The equation below gives an example of a reduction occurring at the surface of the working electrode. Es is the reduction potential of A (if the electrolyte and the electrode are in their standard conditions, then this potential is a standard reduction potential). As E approaches Es, the current on the surface increases, and when E = Es, the concentration of A equals that of the oxidized/reduced A at the surface ([A] = [A]).[4] As the molecules on the surface of the working electrode are oxidized/reduced, they move away from the surface and new molecules come into contact with the surface of the working electrode. The flow of electrons into or out of the electrode causes the current. The current is a direct measure of the rate at which electrons are being exchanged through the electrode-electrolyte interface. When this rate becomes higher than the rate at which the oxidizing or reducing species can diffuse from the bulk of the electrolyte to the surface of the electrode, the current reaches a plateau or exhibits a peak:

Reduction of molecule A at the surface of the working electrode.

The auxiliary and reference electrode work in unison to balance out the charge added or removed by the working electrode. The auxiliary electrode balances the working electrode, but in order to know how much potential it has to add or remove it relies on the reference electrode. The reference electrode has a known reduction potential. The auxiliary electrode tries to keep the reference electrode at a certain reduction potential and to do this it has to balance the working electrode.[5]

Characterization

[edit]

Linear sweep voltammetry can identify unknown species and determine the concentration of solutions. E1/2 can be used to identify the unknown species while the height of the limiting current can determine the concentration. The sensitivity of current changes vs. voltage can be increased by increasing the scan rate. Higher potentials per second result in more oxidation/reduction of a species at the surface of the working electrode.

Variations

[edit]

For reversible reactions cyclic voltammetry can be used to find information about the forward reaction and the reverse reaction. Like linear sweep voltammetry, cyclic voltammetry applies a linear potential over time and at a certain potential the potentiostat will reverse the potential applied and sweep back to the beginning point. Cyclic voltammetry provides information about the oxidation and reduction reactions.

Applications

[edit]

While cyclic voltammetry is applicable to most cases where linear sweep voltammetry is used, there are some instances where linear sweep voltammetry is more useful. In cases where the reaction is irreversible cyclic voltammetry will not give any additional data that linear sweep voltammetry would give us.[6] In one example,[7] linear voltammetry was used to examine direct methane production via a biocathode. Since the production of methane from CO2 is an irreversible reaction, cyclic voltammetry did not present any distinct advantage over linear sweep voltammetry. This group found that the biocathode produced higher current densities than a plain carbon cathode and that methane can be produced from a direct electric current without the need of hydrogen gas.

See also

[edit]

References

[edit]
  1. ^ Douglas A. Skoog; F. James Holler; Stanley R. Crouch (27 January 2017). Principles of Instrumental Analysis. Cengage Learning. pp. 658–. ISBN 978-1-305-57721-3.
  • ^ Nahir, Tal M.; Clark, Rose A.; Bowden, Edmond F. (2002). "Linear-Sweep Voltammetry of Irreversible Electron Transfer in Surface-Confined Species Using the Marcus Theory". Analytical Chemistry. 66 (15): 2595–2598. doi:10.1021/ac00087a027. ISSN 0003-2700.
  • ^ Tissue, Brian M. "Linear Sweep Voltammetry". CHP. Archived from the original on 2013-01-24.
  • ^ "Voltammetry". CHP. Archived from the original on 2003-06-12.
  • ^ Kounaves, Samuel P. Voltammetric Techniques. Handbook of Instrumental Techniques for Analytical Chemistry. pp. 709–725.
  • ^ "Instrumentation, Pine Research. Linear Sweep Voltammetry". CHP. 2008. Archived from the original on 2017-05-20. Retrieved 2013-12-05.
  • ^ Cheng, Shaoan; Xing, Defeng; Call, Douglas F; Logan, Bruce E. (2009). "Direct Biological Conversion of Electrical Current into Methane by Electromethanogenesis". Environ. Sci. Technol. 43 (10): 3953–3958. Bibcode:2009EnST...43.3953C. doi:10.1021/es803531g. PMID 19544913.

  • Retrieved from "https://en.wikipedia.org/w/index.php?title=Linear_sweep_voltammetry&oldid=1190059870"

    Category: 
    Electroanalytical methods
    Hidden categories: 
    Articles with short description
    Short description is different from Wikidata
     



    This page was last edited on 15 December 2023, at 18:23 (UTC).

    Text is available under the Creative Commons Attribution-ShareAlike License 4.0; additional terms may apply. By using this site, you agree to the Terms of Use and Privacy Policy. Wikipedia® is a registered trademark of the Wikimedia Foundation, Inc., a non-profit organization.



    Privacy policy

    About Wikipedia

    Disclaimers

    Contact Wikipedia

    Code of Conduct

    Developers

    Statistics

    Cookie statement

    Mobile view



    Wikimedia Foundation
    Powered by MediaWiki